New optical scheme for a polarimetric-based glucose sensor

J Biomed Opt. 2004 Jan-Feb;9(1):103-15. doi: 10.1117/1.1626664.

Abstract

A new optical scheme to detect glucose concentration in the aqueous humor of the eye is presented. The ultimate aim is to apply this technique in designing a new instrument for, routinely and frequently, noninvasively monitoring blood glucose levels in diabetic patients without contact (no index matching) between the eye and the instrument. The optical scheme exploits the Brewster reflection of circularly polarized light off of the lens of the eye. Theoretically, this reflected linearly polarized light on its way to the detector is expected to rotate its state of polarization, owing to the presence of glucose molecules in the aqueous humor of a patient's eye. An experimental laboratory setup based on this scheme was designed and tested by measuring a range of known concentrations of glucose solutions dissolved in water. (c) 2004 Society of Photo-Optical Instrumentation Engineers.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.
  • Validation Study

MeSH terms

  • Aqueous Humor / chemistry*
  • Blood Glucose / analysis*
  • Computer Simulation
  • Diabetes Mellitus / blood*
  • Diabetes Mellitus / metabolism
  • Equipment Design
  • Equipment Failure Analysis
  • Glucose / analysis*
  • Humans
  • Microscopy, Polarization / instrumentation
  • Microscopy, Polarization / methods
  • Models, Biological*
  • Optics and Photonics / instrumentation*
  • Refraction, Ocular
  • Refractometry / instrumentation*
  • Refractometry / methods*

Substances

  • Blood Glucose
  • Glucose