Development and evaluation of a candidate reference method for the determination of total cortisol in human serum using isotope dilution liquid chromatography/mass spectrometry and liquid chromatography/tandem mass spectrometry

Anal Chem. 2004 Feb 15;76(4):1008-14. doi: 10.1021/ac034966f.

Abstract

Cortisol is an important diagnostic marker for the production of steroid hormones, and accurate measurements of serum cortisol are necessary for proper diagnosis of adrenal function. A candidate reference method involving isotope dilution coupled with liquid chromatography/mass spectrometry (LC/MS) and liquid chromatography/tandem mass spectrometry (LC/MS/MS) has been developed and critically evaluated. An isotopically labeled internal standard, cortisol-d(3), was added to serum, followed by equilibration and solid-phase and ethyl acetate extractions to prepare samples for liquid chromatography/mass spectrometry electrospray ionization (LC/MS-ESI) and liquid chromatography/tandem mass spectrometry electrospray ionization (LC/MS/MS-ESI) analyses. (M + H)(+) ions at m/z 363 and 366 for cortisol and its labeled internal standard were monitored for LC/MS. The transitions of (M + H)(+) --> [(M + H)(+) - 2H(2)O] at m/z 363 --> 327 and 366 --> 330 were monitored for LC/MS/MS. The accuracy of the measurement was evaluated by a comparison of results of this candidate reference method on lyophilized human serum reference materials for cortisol [Certified Reference Materials 192 and 193] with the certified values determined by gas chromatography/mass spectrometry reference methods and by a recovery study for the added cortisol. The results of this method for total cortisol agreed with the certified values within 1.1%. The recovery of the added cortisol ranged from 99.8% to 101.0%. This method was applied to the determination of cortisol in samples of frozen serum pools. Excellent precision was obtained with within-set CVs of 0.3%-1.5% and between-set CVs of 0.04%-0.4% for both LC/MS and LC/MS/MS analyses. The correlation coefficients of all linear regression lines ranged from 0.998 to 1.000. The detection limits (at a signal-to-noise ratio of approximately 3-5) were 10 and 15 pg for LC/MS and LC/MS/MS, respectively. This method, which demonstrates good accuracy and precision, and is free from interferences from structural analogues, qualifies as a candidate reference method and can be used as an alternative reference method to provide an accuracy base to which the routine methods can be compared.

Publication types

  • Comparative Study
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Chromatography, Liquid / methods*
  • Deuterium
  • Evaluation Studies as Topic
  • Humans
  • Hydrocortisone / blood*
  • Quality Control
  • Reference Standards
  • Reference Values
  • Reproducibility of Results
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Tritium

Substances

  • Tritium
  • Deuterium
  • Hydrocortisone