Reducing power line interference in digitised electromyogram recordings by spectrum interpolation

Med Biol Eng Comput. 2004 Jul;42(4):524-31. doi: 10.1007/BF02350994.

Abstract

Interference from power lines (50 or 60 Hz) is the largest source of extraneous noise in many bio-electric signals and is within the bandwidth of many such signals. In this study, two different methods were compared for their efficacy in removing 50 Hz noise added to surface electromyogram (EMG) signals free of power line interference. The first was a simple second-order recursive digital notch filter. The second was an approach called spectrum interpolation, in which it is assumed that the magnitude of the original 50 Hz component of the EMG signal can be approximated by interpolation of the amplitude spectrum of the signal. When the spectrum was based on records containing an integer number of cycles of 50 Hz interference, and the frequency resolution was finer than 1 Hz, spectrum interpolation performed similarly to, or significantly better than, the notch filter (p < 0.01). It was also possible to make spectrum interpolation more robust than the notch filter. The Pearson squared correlation coefficient r2 between clean signals and signals processed using the notch filter was reduced from 0.98 to 0.65 when the interference frequency was increased by 0.5 Hz, but r2 for spectrum interpolation at 0.2 Hz resolution was only reduced from 0.99 to 0.85 if spectral values between approximately 49.5 and 50.5 Hz were modified by interpolation.

MeSH terms

  • Artifacts*
  • Electric Power Supplies*
  • Electromyography / methods*
  • Humans
  • Signal Processing, Computer-Assisted*