Verification of the computational dosimetry system in JAERI (JCDS) for boron neutron capture therapy

Phys Med Biol. 2004 Aug 7;49(15):3353-65. doi: 10.1088/0031-9155/49/15/003.

Abstract

Clinical trials for boron neutron capture therapy (BNCT) by using the medical irradiation facility installed in Japan Research Reactor No. 4 (JRR-4) at Japan Atomic Energy Research Institute (JAERI) have been performed since 1999. To carry out the BNCT procedure based on proper treatment planning and its precise implementation, the JAERI computational dosimetry system (JCDS) which is applicable to dose planning has been developed in JAERI. The aim of this study was to verify the performance of JCDS. The experimental data with a cylindrical water phantom were compared with the calculation results using JCDS. Data of measurements obtained from IOBNCT cases at JRR-4 were also compared with retrospective evaluation data with JCDS. In comparison with phantom experiments, the calculations and the measurements for thermal neutron flux and gamma-ray dose were in a good agreement, except at the surface of the phantom. Against the measurements of clinical cases, the discrepancy of JCDS's calculations was approximately 10%. These basic and clinical verifications demonstrated that JCDS has enough performance for the BNCT dosimetry. Further investigations are recommended for precise dose distribution and faster calculation environment.

Publication types

  • Comparative Study
  • Evaluation Study
  • Validation Study

MeSH terms

  • Algorithms*
  • Body Burden
  • Boron Neutron Capture Therapy / methods*
  • Brain Neoplasms / radiotherapy*
  • Computer Simulation
  • Humans
  • Japan
  • Models, Biological*
  • Quality Assurance, Health Care / methods
  • Radiometry / methods*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity