Determination of microcystins in biological samples by matrix solid-phase dispersion and liquid chromatography-mass spectrometry

J Chromatogr A. 2005 May 6;1073(1-2):257-62. doi: 10.1016/j.chroma.2004.08.128.

Abstract

A method for the detection and quantification of the microcystins (MCs)-MC-LR, MC-RR and MC-YR-in biological samples by matrix solid-phase dispersion (MSPD) has been developed. The optimum extraction conditions were 500 mg of liver or kidney, C18 bonded silica as dispersant, and a mixture methanol-water (70:30) as eluent. The MCs were determined by liquid chromatography electrospray mass spectrometry (LC/ES/MS). Recoveries of biological extracts at three different spiked levels (1-10 mg kg(-1)) ranged from 40.5 to 87.0% in liver, and from 52.5 to 74.5 in kidney. R.S.D.s were < 15.6% and < 10.6%, respectively. The detection and quantification limits were 0.05 and 0.5 mg kg(-1), for all MCs. The method was applied to MCs detection in liver and kidney of rat previously injected i.p. with MC-LR. Results showed the presence of MC-LR in the liver of the animals injected with the highest dose.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Kidney / chemistry
  • Liver / chemistry
  • Male
  • Microcystins
  • Peptides, Cyclic / analysis*
  • Rats
  • Rats, Wistar
  • Reference Standards
  • Reproducibility of Results

Substances

  • Microcystins
  • Peptides, Cyclic
  • microcystin