Influence of experimental parameters on the fluorescence response and recovery of the high-performance liquid chromatography analysis of fumonisin B1

J Chromatogr A. 2006 Mar 24;1109(2):312-6. doi: 10.1016/j.chroma.2006.02.003. Epub 2006 Feb 21.

Abstract

The effect of mobile phase pH, column temperature, extraction method and derivatisation time on the fluorescence response and recovery of fumonisin B1 in corn was investigated. Column temperature and mobile phase pH were negatively and positively correlated with the fluorescence response, respectively. Use of an Ultraturrax blender for extraction resulted in higher FB1 recoveries compared to a rotary shaker. In contrast to other reports, maximum fluorescence response occurred after a derivatisation time of 8 min. These results reflect on the absolute requirement for standardisation of the aforementioned parameters.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Fumonisins / analysis*
  • Hydrogen-Ion Concentration
  • Reference Standards
  • Spectrometry, Fluorescence / methods*
  • Temperature

Substances

  • Fumonisins
  • fumonisin B1