Molecularly imprinted polymer diffraction grating as label-free optical bio(mimetic)sensor

Biosens Bioelectron. 2011 Jan 15;26(5):2801-4. doi: 10.1016/j.bios.2010.11.009. Epub 2010 Dec 4.

Abstract

Micropatterned molecularly imprinted polymer (MIP) transmissive 2D diffraction gratings (DGs) are fabricated and evaluated as label-free antibiotic bio(mimetic)sensors. Polymeric gratings are prepared by using microtransfer molding based on SiO(2)/Si molds. The morphology of the MIP gratings is studied by optical and atomic force microscopes. MIP 2D-DGs exhibit 2D optical diffraction patterns, and measurement of changes in diffraction efficiency is used as sensor response. The refractive index of the micropatterned MIP material was estimated, via solvent index matching experiments, to be 1.486. Immersion of a MIP 2D-DG in different solutions of target-antibiotic enrofloxacin leads to significant variations in diffraction efficiency, demonstrating target-molecule detection. On the other hand, no significant response is observed for both control experiments: MIP grating exposed to a non-retained analyte and an equivalent non-imprinted polymer grating exposed to the target analyte, showing highly specific antibiotic label-free optical recognition.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anti-Bacterial Agents / analysis*
  • Biomimetic Materials*
  • Biosensing Techniques / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis
  • Molecular Probe Techniques*
  • Refractometry / instrumentation*
  • Staining and Labeling

Substances

  • Anti-Bacterial Agents