Electrospray Ion Chromatography-Tandem Mass Spectrometry of Bromate at Sub-ppb Levels in Water

Anal Chem. 1996 Aug 1;68(15):2554-8. doi: 10.1021/ac960175k.

Abstract

An electrospray ion chromatography-tandem mass spectrometry (IC-MS/MS) method has been developed for the analysis of bromate ions in water. This IC-MS/MS method improves the limit of detection of bromate ions by a factor of 10. The method consists of solid phase extraction with an ion exchange column and elution of the analyte with water/methanol ammonium sulfate eluent on-line with a negative ion electrospray mass spectrometry detection. SPE requires sample pretreatment to remove any major ions that displace bromate, consisting of eliminating SO(4)(2)(-), Cl(-), and HCO(3)(-) ions respectively with barium-form, silver-form, and acid (H(+)-form) exchange resins. The methanolic sulfate eluent permits IC-MS coupling via an electrospray interface. BrO(3)(-) was selected in the first quadrupole (Q1) at two m/z values, 127 and 129, according to the isotope contributions of (79)Br and (81)Br. After fragmentation in the collision cell (second quadrupole, Q2), the third quadrupole (Q3) analyzes the product ions as (M - O)(-), (M - 2O)(-), and (M - 3O)(-). Among the six recordable transitions, four were selected, the other two yielding high background. A lowered resolution raised sensitivity by a factor of up to 3. The limit of quantitation of this method was 0.1 μg/L.