Characterizing the inflammatory reaction in explanted Medtronic Freestyle stentless porcine aortic bioprosthesis over a 6-year period

Cardiovasc Pathol. 2012 May-Jun;21(3):158-68. doi: 10.1016/j.carpath.2011.05.003. Epub 2011 Aug 3.

Abstract

Background: The Medtronic Freestyle valve is a stentless porcine valve with reportedly excellent clinical and hemodynamic results, but little has been reported about its long-term pathology.

Methods: Seventeen Freestyle valves were explanted (from 2003 to 2009) and reviewed to assess reasons for bioprosthesis failure. All valves were examined in detail, using histochemistry and immunohistochemistry to identify morphological changes, as well as cellular and humoral responses.

Results: One Freestyle valve, explanted for mitral valve endocarditis on the fifth postoperative day, was excluded from analysis. The average implant duration was 71.1±35.2 months. Six valves were explanted for infective endocarditis, six for aortic insufficiency, and four for aortic stenosis. Calcification was seen in 11 explants, pannus in 15, thrombus in 12, cusp tears in 9, and 10 explants showed needle tract-like injuries. A chronic inflammatory reaction involving the xenograft arterial wall was seen in 15 of 16 valves. The cells were composed of macrophages and lymphocytes, including T cells (CD8 positive) and B cells. Significant damage to the porcine aortic wall was seen in 15 cases, and cusp myocardial shelf damage in 7 cases. All cases stained positively for IgG and C4dpar.

Conclusions: The porcine aortic tissue showed T cell-mediated rejection and significant aortic medial damage, consistent with dilatation of the porcine aortic root. The demonstration of IgG suggests the likelihood of humoral rejection, in addition to cellular rejection. One of the underlying possibilities is that the porcine aortic tissues are inadequately fixed, hence the retained antigenicity.

MeSH terms

  • Adult
  • Aged
  • Animals
  • Aortic Valve / pathology
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / pathology
  • Aortic Valve Insufficiency / surgery
  • Aortic Valve Stenosis / pathology
  • Aortic Valve Stenosis / surgery
  • Bioprosthesis*
  • Device Removal
  • Endocarditis / pathology
  • Endocarditis / surgery
  • Equipment Failure Analysis*
  • Female
  • Fixatives
  • Graft Rejection
  • Heart Valve Prosthesis Implantation / adverse effects*
  • Heart Valve Prosthesis*
  • Humans
  • Inflammation / etiology*
  • Inflammation / pathology
  • Male
  • Middle Aged
  • Prosthesis Failure
  • Prosthesis-Related Infections / etiology*
  • Prosthesis-Related Infections / pathology
  • Swine

Substances

  • Fixatives