Evaluation of composite mesh for ventral hernia repair

JSLS. 2011 Jul-Sep;15(3):298-304. doi: 10.4293/108680811X13071180407393.

Abstract

Introduction: Composite mesh prostheses incorporate the properties of multiple materials for ventral hernia repair. This study evaluated a polypropylene/ePTFE composite mesh with a novel internal polydioxanone (PDO) absorbable ring.

Methods: Composite mesh was placed intraperitoneally in 16 pigs through an open laparotomy and explanted at 2, 4, 8, and 12 weeks. Intraabdominal adhesions were measured laparoscopically. Host tissue in-growth was assessed histologically and tensiometrically. Degradation of the internal PDO ring component was also measured tensiometrically. Appropriate statistical tests were used, and P ≤.05 indicated significance.

Results: No adhesions were formed in 50% of the grafts explanted at 8 weeks and 25% of grafts explanted at 12 weeks. There were significantly more vascular structures at 8 weeks, 73.5 ± 28, compared with 2 weeks, 6.75 ± 2 (P ≤.01). The T-peel force at the mesh-host tissue interface was not significantly different among time points. The absorbable PDO ring underwent complete degradation by 12 weeks.

Conclusions: This composite mesh was associated with minimal intraabdominal adhesions, progressive in-growth of host tissues, and complete degradation of a novel internal PDO ring that aided mesh positioning. This composite hernia mesh showed a favorable performance in a porcine model of open ventral hernia repair.

MeSH terms

  • Animals
  • Equipment Design
  • Female
  • Hernia, Ventral / surgery*
  • Polydioxanone
  • Polypropylenes
  • Polytetrafluoroethylene
  • Surgical Mesh*
  • Swine

Substances

  • Polypropylenes
  • Polydioxanone
  • Polytetrafluoroethylene