Noncontact ECG recording system with real time capacitance measurement for motion artifact reduction

IEEE Trans Biomed Circuits Syst. 2014 Oct;8(5):617-25. doi: 10.1109/TBCAS.2014.2359053. Epub 2014 Oct 13.

Abstract

A system for noncontact ECG recording is proposed that measures the real time electrode-body capacitance concurrently with the ECG as a reference signal for motion artifact reduction. Simultaneous recordings of these two signals from the human body in the presence of electrode motion artifacts are shown and an adaptive least-mean-squares (LMS) filtering algorithm run on these signals is demonstrated to be able to reduce the severity of certain types of electrode motion artifacts.

MeSH terms

  • Algorithms
  • Artifacts*
  • Electrocardiography / instrumentation*
  • Electrocardiography / methods
  • Equipment Design
  • Humans
  • Movement
  • Signal Processing, Computer-Assisted / instrumentation*