Monitoring of positron using high-energy gamma camera for proton therapy

Ann Nucl Med. 2015 Apr;29(3):268-75. doi: 10.1007/s12149-014-0936-4. Epub 2014 Dec 5.

Abstract

Purpose: In proton therapy, imaging of proton-induced positrons is a useful method to monitor the proton beam distribution after therapy. Usually, a positron emission tomography (PET) system installed in or near the proton beam treatment room is used for this purpose. However, a PET system is sometimes too large and expensive for this purpose. We developed a small field-of-view (FOV) gamma camera for high-energy gamma photons and used it for monitoring the proton-induced positron distribution.

Methods: The gamma camera used 0.85 mm × 0.85 mm × 10 mm Ce:Gd3Al2Ga3O12 (GAGG) pixels arranged in 20 × 20 matrix to form a scintillator block, which was optically coupled to a 1-inch-square position-sensitive photomultiplier tube (PSPMT). The GAGG detector was encased in a 20-mm thick container and a pinhole collimator was mounted on its front. The gamma camera was set 1.2 m from the 35 cm × 35 cm × 5 cm plastic phantom in the proton therapy treatment room, and proton beams were irradiated to the phantom with two proton energies.

Results: The gamma camera had spatial resolution of ~6.7 cm and sensitivity of 3.2 × 10(-7) at 1 m from the collimator surface. For both proton energies, positron distribution in the phantom could be imaged by the gamma camera with 10-min acquisition. The lengths of the range of protons measured from the images were almost identical to the simulation results.

Conclusions: These results indicate that the developed high-energy gamma camera is useful for imaging positron distributions in proton therapy.

Publication types

  • Evaluation Study

MeSH terms

  • Electrons
  • Equipment Design
  • Gamma Cameras*
  • Humans
  • Phantoms, Imaging
  • Proton Therapy / instrumentation*
  • Proton Therapy / methods*
  • Sodium Radioisotopes