Valence EELS below the limit of inelastic delocalization using conical dark field EFTEM or Bessel beams

Ultramicroscopy. 2017 Feb:173:24-30. doi: 10.1016/j.ultramic.2016.11.022. Epub 2016 Nov 22.

Abstract

In this experimental work we present novel methods to increase the spatial resolution of valence electron energy loss spectrometry (VEELS) investigations below the limit given by the inelastic delocalization. For this purpose we analyse a layer stack consisting of silicon/silicon-oxide/silicon-nitride/silicon-oxide/silicon (SONOS) with varying layer thickness down to the 2nm level. Using a combination of a conical illumination and energy filtered transmission electron microscopy we are able to identify the layers by using low energy losses. Employing Bessel beams we demonstrate that VEELS can be performed in dark-field conditions while simultaneously the Bessel beam increases the spatial resolution of the elastic image due to less sensitivity to the spherical aberration of the condenser lens system. The dark-field conditions also guarantee that only electrons are collected that have neither undergone an energy loss being due to the Cˇerenkov effect, nor due to the excitation of transition radiation or light guiding modes. We consequently are able to measure the optical properties of a 2.5nm thin oxide being sandwiched by the silicon substrate and a silicon-nitride layer.

Keywords: Bessel beams; Coulomb interaction; Inelastic delocalization; VEELS.