Multiple-reaction monitoring for multiplex detection of three bacterial toxins using liquid chromatography-tandem mass spectrometry

Lett Appl Microbiol. 2017 Mar;64(3):217-224. doi: 10.1111/lam.12706. Epub 2017 Jan 30.

Abstract

Clostridium perfringens epsilon toxin, staphylococcal enterotoxin B and shiga toxin are implicated in a number of diseases and food-borne intoxications and are considered potential agents for bioterrorism and warfare. Artificially generated aerosol is the likely mode of delivery of these for nefarious uses, potentially capable of causing mass destruction to human and animal health by inhalation of toxic bioaerosol. Multiplex and unambiguous detection of these agents is of paramount importance for emergency response in a biothreat scenario and for food safety. Multiple-reaction monitoring (MRM) assay for simultaneous monitoring of the three toxins is reported here using reverse-phase high-performance liquid chromatography-electrospray ionization-tandem mass spectrometry. Three different peptides with two fragment ions each were considered for quantification and confirmation. One of the three MRM transitions from each toxin, which exhibited the best sensitivity, was selected for multiplexing of the assay. Simulating a biothreat scenario wherein the bioaerosol is collected in 10 ml of buffer, the multiplex assay was tested with blind samples with one or more of the three toxins even in the presence of interfering Escherichia coli lysate proteins.

Keywords: MRM; bacterial toxins; bioterrorism; mass spectrometry; multiplex detection.

Publication types

  • Evaluation Study

MeSH terms

  • Bacterial Infections / microbiology
  • Bacterial Toxins / analysis*
  • Chromatography, High Pressure Liquid / methods*
  • Enterotoxins / analysis*
  • Humans
  • Peptides / chemistry
  • Tandem Mass Spectrometry / methods*

Substances

  • Bacterial Toxins
  • Clostridium perfringens epsilon-toxin
  • Enterotoxins
  • Peptides
  • enterotoxin B, staphylococcal