Respiratory effort from the photoplethysmogram

Med Eng Phys. 2017 Mar:41:9-18. doi: 10.1016/j.medengphy.2016.12.010. Epub 2017 Jan 23.

Abstract

The potential for a simple, non-invasive measure of respiratory effort based on the pulse oximeter signal - the photoplethysmogram or 'pleth' - was investigated in a pilot study. Several parameters were developed based on a variety of manifestations of respiratory effort in the signal, including modulation changes in amplitude, baseline, frequency and pulse transit times, as well as distinct baseline signal shifts. Thirteen candidate parameters were investigated using data from healthy volunteers. Each volunteer underwent a series of controlled respiratory effort maneuvers at various set flow resistances and respiratory rates. Six oximeter probes were tested at various body sites. In all, over three thousand pleth-based effort-airway pressure (EP) curves were generated across the various airway constrictions, respiratory efforts, respiratory rates, subjects, probe sites, and the candidate parameters considered. Regression analysis was performed to determine the existence of positive monotonic relationships between the respiratory effort parameters and resulting airway pressures. Six of the candidate parameters investigated exhibited a distinct positive relationship (p<0.001 across all probes tested) with increasing upper airway pressure repeatable across the range of respiratory rates and flow constrictions studied. These were: the three fundamental modulations in amplitude (AM-Effort), baseline (BM-Effort) and respiratory sinus arrhythmia (RSA-Effort); two pulse transit time modulations - one using a pulse oximeter probe and an ECG (P2E-Effort) and the other using two pulse oximeter probes placed at different peripheral body sites (P2-Effort); and baseline shifts in heart rate, (BL-HR-Effort). In conclusion, a clear monotonic relationship was found between several pleth-based parameters and imposed respiratory loadings at the mouth across a range of respiratory rates and flow constrictions. The results suggest that the pleth may provide a measure of changing upper airway dynamics indicative of the effort to breathe.

Keywords: Clinical monitoring; Photoplethysmogram; Pulse amplitude modulation; Pulse oximetry; Pulse transit time; Respiratory effort; Respiratory sinus arrhythmia.

MeSH terms

  • Adult
  • Algorithms
  • Female
  • Humans
  • Male
  • Photoplethysmography*
  • Pilot Projects
  • Pressure
  • Respiration*
  • Young Adult