Because reactive oxygen species are involved in a range of pathologies, developing analytical tools for this group of molecules opens new vistas for biomedical diagnostics. Herein, we fabricate a porous silicon microcavity (pSiMC) functionalized with luminescent singlet oxygen (1O2) probe EuA ((Eu(III)-2,2',2″-(10-(2-((4-(2-((4-(2-((anthracen-9-ylmethyl)amino)ethyl)-1H-1,2,3-triazol-1-yl)amino)-2-oxoethyl)-2-oxo-1,2-dihydroquinolin-7-yl)amino)-2-oxoethyl)-1,4,7,10-tetraazacyclododecane-1,4,7-triyl)triacetic acid) as proof of concept of an optical sensor for reactive oxygen species. We characterize each surface modification step of the pSiMC by means of FTIR and X-ray photoelectron spectroscopy as well as by determining the optical shifts of the resonance wavelength of the pSiMC. The luminescence signal upon detection of 1O2 on the EuA-modified pSiMC is enhanced ∼2-fold compared to that of a single layer and a detuned microcavity. The sensing performance of the EuA probe is improved significantly on the pSiMC compared to that in aqueous solution, giving a limit of 1O2 detection of 3.7 × 10-8 M.