Quantitative two-dimensional ultrashort echo time magnetization transfer (2D UTE-MT) imaging of cortical bone

Magn Reson Med. 2018 Apr;79(4):1941-1949. doi: 10.1002/mrm.26846. Epub 2017 Aug 3.

Abstract

Purpose: To investigate quantitative 2D ultrashort echo time magnetization transfer (UTE-MT) imaging in ex vivo bovine cortical bone and in vivo human tibial cortical bone.

Methods: Data were acquired from five fresh bovine cortical bone samples and five healthy volunteer tibial cortical bones using a 2D UTE-MT sequence on a clinical 3T scanner. The 2D UTE-MT sequence used four or five MT powers with five frequency offsets. Results were analyzed with a two-pool quantitative MT model, providing measurements of macromolecular fraction (f), macromolecular proton transverse relaxation times (T2m ), proton exchange rates from water/macromolecular to the macromolecular/water pool (RM0m /RM0w ), and spin-lattice relaxation rate of water pool (R1w ). A sequential air-drying study for a small bovine cortical bone chip was used to investigate whether above MT modeling parameters were sensitive to the water loss.

Results: Mean fresh bovine cortical bone values for f, T2m , R1w , RM0m , and RM0w were 59.9 ± 7.3%, 14.6 ± 0.3 μs, 9.9 ± 2.4 s-1 , 17.9 ± 3.6 s-1 , and 11.8 ± 2.0 s-1 , respectively. Mean in vivo human cortical bone values for f, T2m , R1w , RM0m and RM0w were 54.5 ± 4.9%, 15.4 ± 0.6 μs, 8.9 ± 1.1 s-1 , 11.5 ± 3.5 s-1 , and 9.5 ± 1.9 s-1 , respectively. The sequential air-drying study shows that f, RM0m , and R1w were increased with longer drying time.

Conclusion: UTE-MT two-pool modeling provides novel and useful quantitative information for cortical bone. Magn Reson Med 79:1941-1949, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Keywords: cortical bone; magnetization transfer; two-pool modeling; ultrashort echo time.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Algorithms
  • Animals
  • Cattle
  • Cortical Bone / diagnostic imaging*
  • Echo-Planar Imaging*
  • Humans
  • Image Enhancement
  • Image Interpretation, Computer-Assisted
  • Male
  • Osteoporotic Fractures / diagnostic imaging
  • Protons
  • Reproducibility of Results
  • Tibia / diagnostic imaging*

Substances

  • Protons