Improving the quality of preclinical research echocardiography: observations, training, and guidelines for measurement

Am J Physiol Heart Circ Physiol. 2018 Jul 1;315(1):H58-H70. doi: 10.1152/ajpheart.00157.2018. Epub 2018 Apr 20.

Abstract

Informal training in preclinical research may be a contributor to the poor reproducibility of preclinical cardiology research and low rates of translation into clinical research and practice. Mouse echocardiography is a widely used technique to assess cardiac structure and function in drug intervention studies using disease models. The interobserver variability of clinical echocardiographic measurements has been shown to improve with formalized training, but preclinical echocardiography lacks similarly critical standardization of training. The aims of this investigation were to assess the interobserver variability of echocardiographic measurements from studies in mice and address any technical impediments to reproducibility by implementing standardized guidelines through formalized training. In this prospective, single-site, observational cohort study, 13 scientists performing preclinical echocardiographic image analysis were assessed for measurement of short-axis M-mode-derived dimensions and calculated left ventricular (LV) mass. Ten M-mode images of mouse hearts acquired and analyzed by an expert researcher with a spectrum of LV mass were selected for assessment and validated by autopsy weight. After the initial observation, a structured formal training program was introduced, and accuracy and reproducibility were reevaluated. Mean absolute percentage error for expert-calculated LV mass was 6 ± 4% compared with autopsy LV mass and 25 ± 21% for participants before training. Standardized formal training improved participant mean absolute percentage error by ~30% relative to expert-calculated LV mass ( P < 0.001). Participants initially categorized with high-range error (25-45%) improved to low-moderate error ranges (<15-25%). This report reveals an example of technical skill training insufficiency likely endemic to preclinical research and provides validated guidelines for echocardiographic measurement for adaptation to formalized in-training programs. NEW & NOTEWORTHY The informal training common to academic/research institutions may be a contributor to the relatively poor reproducibility observed for preclinical cardiac research. In our observation of echocardiography analysis in murine models, we present evidence of moderate interobserver variability in standard preclinical research practice at an Australian heart research institute. These observations give rise to our recommendations for practical guidelines for echocardiography analysis in an adaptable approach to general preclinical research skill training. Listen to this article's corresponding podcast at https://ajpheart.podbean.com/e/preclinical-echocardiography-training-and-guidelines/ .

Keywords: echocardiography; preclinical research; reproducibility; research skills training; variability.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Data Accuracy
  • Echocardiography / methods
  • Echocardiography / standards*
  • Humans
  • Practice Guidelines as Topic*
  • Reproducibility of Results
  • Translational Research, Biomedical / education
  • Translational Research, Biomedical / standards*