A liquid chromatography-tandem mass spectrometry (LC-MS/MS)-based assay for simultaneous quantification of aldosterone, renin activity, and angiotensin II in human plasma

J Chromatogr B Analyt Technol Biomed Life Sci. 2021 Aug 1:1179:122740. doi: 10.1016/j.jchromb.2021.122740. Epub 2021 Jul 3.

Abstract

Accurate quantification of plasma aldosterone (ALD) and renin activity (PRA)is critical for the diagnosis of primary aldosteronism (PA). Liquid chromatography-tandem mass spectrometry (LC-MS/MS) is considered the "gold standard" method for the determination of ALDand PRA. The aim of this study is to develop a new LC-MRM/MS assay for quantifying plasma ALD, PRA, and angiotensin II (Ang II) simultaneously and validate its effectiveness. To be more specific, plasmasamples were prepared by solid-phase extraction and separated in an ultra-performance reversed-phase column. MS detection was performed via a triple quadrupole mass spectrometer containing both positive and negative ion monitoring modes. The developed assay was then validated according to the standard guidelines and the influence of sample incubation on ALD and Ang II concentration was evaluated. In addition, the variation of endogenous Ang I was explored. The proposed LC-MRM/MS method was compared another LC-MS/MS method, which detects ALD, Ang I, and Ang II separately. Analyteswere separated and quantified within 5 min. The assay wasvalidated to be linear up to 5000 pg/ml for ALD and Ang II and 33.3 ng/ml/h for PRA.The lower limit of quantification (LLOQ) was 15 pg/ml, 15 pg/ml, and 0.1 ng/ml/hfor ALD, Ang II, and PRArespectively. Specificity, precision, accuracy, and stability were tested to meet the requirements of the guidelines. Significant changes were not found in ALD and Ang II concentrations over the 3 h-incubation. In addition, it was demonstratedthat the resultof PRA was not stronglyinfluenced by the endogenous Ang I. Comparison with another LC-MS/MS method was performed using the same apparatusand the proposed method was proved to be in good coincidence with the correlation coefficients rangingfrom 0.955to0.996. A sensitive and reliable method for simultaneousquantification of ALD, PRA, and Ang II has been developed and this study will significantly promote laboratory workflow efficiency and throughput.

Keywords: Aldosterone; Angiotensin II; Liquid chromatography-tandem mass spectrometry; Renin activity.

MeSH terms

  • Aldosterone / blood*
  • Angiotensin II / blood*
  • Chromatography, Liquid / methods*
  • Humans
  • Limit of Detection
  • Linear Models
  • Renin / blood*
  • Renin / metabolism
  • Reproducibility of Results
  • Tandem Mass Spectrometry / methods*

Substances

  • Angiotensin II
  • Aldosterone