Conventional Audiometry, Extended High-Frequency Audiometry, and DPOAE for Early Diagnosis of NIHL

Iran Red Crescent Med J. 2014 Jan;16(1):e9628. doi: 10.5812/ircmj.9628. Epub 2014 Jan 5.

Abstract

Background: Noise most frequently affects hearing system, as it may typically cause a bilateral, progressive sensorineural hearing loss at high frequencies.

Objectives: This study was designed to compare three different methods to evaluate noise-induced hearing loss (conventional audiometry, high-frequency audiometry, and distortion product otoacoustic emission).

Material and methods: This was a cross-sectional study. Data was analyzed by SPSS (ver. 19) using chi square, T test and repeated measures analysis. Study samples were workers from tile and ceramic industry.

Results: We found that conventional audiometry, extended high-frequency audiometry, low-tone distortion product otoacoustic emission and high-tone distortion product otoacoustic emission had abnormal findings in 29 %, 69 %, 22 %, and 52 % of participants. Most frequently affected frequencies were 4000 and 6000Hz in conventional audiometry, and 14000 and 16000 in extended high-frequency audiometry.

Conclusions: Extended high-frequency audiometry was the most sensitive test for detection of hearing loss in workers exposed to hazardous noise compared with conventional audiometry and distortion product otoacoustic.

Keywords: Audiometry; Hearing Loss; Noise; Otoacoustic Emissions, Spontaneous.