Establishment of relationships between native and inhalation device specific spirometric parameters as a step towards patient tailored inhalation device selection

Respir Med. 2019 Jul-Aug:154:133-140. doi: 10.1016/j.rmed.2019.06.021. Epub 2019 Jun 22.

Abstract

Introduction: Drug emission from DPIs is dependent on the inspiratory flow parameters through them, which are not directly measured by standard spirometry. Their estimation based on native spirometric data could help in choosing the appropriate device and optimizing the drug deposition.

Objectives: The aim of this study was to survey patient preferences and to find correlations between breathing parameters of COPD patients through DPI devices and their baseline spirometric data, age, gender, disease severity and anthropometric characteristics. Another objective was to establish relationships between peak inspiratory flows (PIFdev) through Breezhaler®, Genuair® and Turbuhaler® inhalers and their determinants.

Methods: Breathing parameters of 49 patients with previously diagnosed COPD and currently using one of the above inhalers were recorded by normal spirometry and while inhaling through the selected DPIs. Statistical analysis of the measured data was completed. All specific data are provided as (mean ± standard deviation).

Results: More than 60% of the patients stated that their current device is the easiest to use. The means of the measured PIFdev values were 91.4 L/min, 77.1 L/min and 77.5 L/min for Breezhaler®, Genuair®, and Turbuhaler®, respectively. PIFdev values were significantly higher for males than for females, but differences upon age, BMI and disease severity group were not significant (at p = 0.05). Peak inspiratory flows through the inhalers (PIFdev) correlated best with their native spirometric counterparts (PIF) and linear PIFdev-PIF relationships could be determined (Breezhaler®: r = 0.60, p = 0.002, Genuair®: r = 0.55, p = 0.001, Turbuhaler®: r = 0.57, p = 0.002). Physical background of the deduced equations was also provided.

Conclusions: Present correlations may be used to assess the success of inhalation of COPD patients through the studied devices and to choose the appropriate device for each patient. As a consequence, the amount of the drug emitted by the device can be optimized, the deposition efficiency within the lungs increased and the related therapeutic effect improved.

Keywords: Management of COPD patients; Patient and device specific spirometric parameters; Personalized device selection.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Administration, Inhalation
  • Aged
  • Algorithms
  • Anthropometry / methods
  • Dry Powder Inhalers / instrumentation*
  • Equipment Design / statistics & numerical data
  • Equipment Design / trends
  • Female
  • Humans
  • Hungary / epidemiology
  • Inspiratory Capacity / drug effects
  • Inspiratory Capacity / physiology
  • Male
  • Middle Aged
  • Patient Care Management / methods
  • Patient Preference / statistics & numerical data*
  • Pulmonary Disease, Chronic Obstructive / drug therapy*
  • Pulmonary Disease, Chronic Obstructive / physiopathology
  • Pulmonary Disease, Chronic Obstructive / psychology
  • Severity of Illness Index
  • Spirometry / statistics & numerical data*
  • Surveys and Questionnaires